Practical Application of a New Cuffless Blood Pressure Measurement Method.
Gogiberidze N, Suvorov A, Sultygova E, Sagirova Z, Kuznetsova N, Gognieva D, Chomakhidze P, Frolov V, Bykova A, Mesitskaya D, Novikova A, Kondakov D, Volovchenko A, Omboni S, Kopylov P.
Gogiberidze N, et al.
Pathophysiology. 2023 Dec 1;30(4):586-598. doi: 10.3390/pathophysiology30040042.
Pathophysiology. 2023.
PMID: 38133143
Free PMC article.
In the subgroups, the smallest error (less than 3 mm Hg) was observed in the prehypertension group, with a slightly larger error (up to 4 mm Hg) found among patients with a normal blood pressure and stage 1 hypertension. The largest error was found in the stage 2 hy …
In the subgroups, the smallest error (less than 3 mm Hg) was observed in the prehypertension group, with a slightly larger error (up to 4 mm …